Easy to use, hand-held Thermocouple indicator with 2 input channels. Simply connect your thermocouple sensor(s) to the device using a miniature flat-pinned plug. Selectable between thermocouple inputs K, J, T, N, E, R and S.

Brand image

Easy to use, hand-held Thermocouple indicator with 2 input channels. Simply connect your thermocouple sensor(s) to the device using a miniature flat-pinned plug. Selectable between thermocouple inputs K, J, T, N, E, R and S.

The indicator has a sampling rate of 1.5 seconds and can show both input channels on the screen at once, identified as T1, T2. A memory function allows storage and recall of up to 44 data snapshots. An alarm can be configured to audibly alert if the temperature exceeds a high/low limit.

Robust construction with an easy-to-read, backlit LCD display. The strong plastic case has an integrated tilt stand. Supplied with instruction manual and batteries. A removable rubber holster is available separately.

Features

  • 2 Thermocouple input channels (T1-T2)
  • Thermocouple Types K, J, T, N, E, R, S
  • °C / °F / K selectable
  • Backlit 4 1/2 digit LCD display
  • MIN, MAX, AVG, HOLD, MEM (44 slots), Hi/Lo Alarm
  • Battery powered, auto power off, low battery indicator
  • Integrated tilt stand
  • Supplied with instruction manual and batteries
  • Rubber protective holster also available
More Information
Product Extra Information

Specification

Input (miniature plug) 1 2 x Thermocouple
Measurement Range K -150°C to +1370°C (-200°F to +1999.9°F)
J -150°C to +1090°C (-200°F to +1994°F)
T -150°C to +400°C (-200°F to +752°F)
N -150°C to +1300°C (-200°F to +1999.9°F)
E -150°C to +870°C (-200°F to +1598°F)
R, S +2°C to +1767°C (+35°F to +1999.9°F)
Accuracy 2 K, J, T, N, E ±[0.05% of reading +0.5°C (0.9°F)]. Below -100°C (-148°F): add 0.15% of reading for K, J, N, E; add 0.45% of reading for T
R, S ±[0.05% of reading +2°C (4°F)]
Temperature Coefficient 0.01% of reading +0.03°C per °C outside specified +18°C ~ +28°C range. Below -100°C: add 0.04% of reading for K, J, N, E; add 0.08% of rdg for T
Sample Rate Every 1.5 seconds
Display Backlit 4 1/2 digit LCD display
Functions Min, Max, Avg, Hold, Mem, Auto Off, Low Battery
Operating Temperature/Humidity 0°C to +50°C (+32°F to +122°F) below 80 %RH
Storage Temperature/Humidity -10°C to +60°C (+14°F to +140°F) below 70 %RH
Power 6 x 1.5V AAA batteries (supplied)
Dimensions (mm) (H x W x D) 155 x 77 x 37 (wih holster); 150 x 72 x 35 (without holster)
Weight (with batteries) ~291g (with holster); ~ 239g (without holster)

1 Thermocouple sensors and miniature plugs not supplied - available separately

2 Accuracy stated is for indicator only. The thermocouple sensor innacuracy should also be considered