Compact, easy-to-use heat source for checking the accuracy of temperature sensors up to +300°C (+572°F) on site, without the need for heavy, expensive equipment. Simply plug in, switch on, set the heat source to the desired temperature test point and insert the test probe into the correct sized well to get a verification of the performance.
The HC-300DW is controlled by a PID temperature controller system incorporating a heater and precision platinum RTD sensor housed in the aluminium block. Fan cooling allows rapid changes in block temperature upon demand. The required temperature may be adjusted in 0.1°C increments using the buttons on the front panel.
Supplied with instruction manual, power cable and in-house test certificate.
Features Easily calibrate thermocouples, RTDs, bi-metals and general thermometers up to +300°C
2 x 3.5mm, 4.2mm, 5.0mm, 6.8mm test hole diameters
100mm well length (90mm insertion depth)
Fast heating/cooling time
Compact size
Overload alarm LEDs
Technical Details |
---|
Temperature Range | +33°C to +300°C (+91.4°F to +572°F) at +20°C ambient |
Resolution | 0.1°C/°F |
Accuracy | ±0.8°C (up to +100°C), ±1.6°F (up to +212°F) |
±1.6°C (+100°C to +200°C), ±3.2°F (+212°F to +392°F) |
±2.8°C (+200°C to +300°C), ±5.6°F (+392°F to +572°F) |
Stability | ±0.1°C (+33°C to +100°C), ±0.2°F (+91.4°F to +212°F) |
±0.2°C (+100°C to +200°C), ±0.4°F (+212°F to +392°F) |
±0.4°C (+200°C to +300°C), ±0.8°F (+392°F to +572°F) |
Test Hole Diameters | 2 x 3.5mm, 4.2mm, 5.0mm, 6.8mm |
Well Length | 100mm (90mm insertion depth) |
Heating Time | 30 minutes to maximum |
Cooling Time | 30 minutes to +100°C (+212°F) |
Stabilization | 5 minutes |
Power | 230 Volts (115 Volts also available) |
Dimensions | 113mm (H) x 183mm (W) x 235mm (D) |
Weight | ~ 3kg |